Guralp Systems CMG-DM24 User Manual
Page 35

Operator’s Guide DM-24 Digitiser
Issue G January 2003
33
3.2 DIGITISER CONFIGURATION:– USING ‘TERMINAL’
DIGITISER CONFIGURATION
Command
Syntax
Comments
SET-ID
System Identifier ? (e.g.
ALPHA,)
Serial # ? (e.g. 1234,00)
interactive
NUNAME,
4339,00
Both sets of characters
must include the comma in
the location shown
SENSOR-TYPE
n SENSOR-TYPE
n=1 CMG-40T
n=2 CMG-3ESP
n=3 CMG-3T
n=4 CMG-3TD
GPS-TYPE
n GPS-TYPE
n = 0 no time signal
n = 1Trimble
n = 2 Garmin
n = 3 stream-sync
Not all versions support all
combinations
NOTE: The last two digits of the serial number should set to 00 (two zeros) as the
system replaces these with the characters to identify the component (Z, N, E etc) and