AMETEK SLM Series Rev B User Manual
Page 48
GPIB/RS-232 Programming Operation
SLM-Series AC/DC Electronic Load
4-6
M540072-01 Rev B
COMMAND NOTE
RETURN
[SYStem:] CHANnel{SP}{1|2|3|4}[A|B]{;|NL}
[SYStem:] CHANnel{SP}{?}{;|NL}
{1|2|3|4}[A|B]
[SYStem:] RECall{SP}{m[,n]}{;|NL}
M=1~5 n=1~30
[SYStem:] STORe{SP}{m[,n]}{;|NL}
M=1~5 n=1~30
[SYStem:] REMOTE{;|NL}
Only RS232 cmd
[SYStem:] LOCAL{;|NL}
Only RS232 cmd
0=:OFF, 1=:ON
[SYStem:] NAME{?}{;|NL}
“XXXXX”
Table 4-5 System Commands - All Modules
COMMAND
SLM DC
SLD
SLM AC
RETURN
MEASure:CURRent {?}{;
⎢NL}
•
•
•
###.####
MEASure:VOLTage {?}{;
⎢NL}
•
•
•
###.####
MEASure:PWR {?}{;
⎢NL}
•
###.####
MEASure:VA {?}{;
⎢NL}
•
###.####
LIM:CURRent:{HIGH|LOW
}{SP}{NR2}{;|NL}
•
•
•
LIM:CURRent:{HIGH|LOW
}{?}{;|NL}
•
•
•
###.####
LIM:POWer:{HIGH|LOW
}{SP}{NR2}{;|NL}
•
•
LIM:POWer:{HIGH|LOW
}{?}{;|NL}
•
•
###.####
LIM:VA:{HIGH|LOW
}{SP}{NR2}{;|NL}
•
LIM:VA:{HIGH|LOW
}{?}{;|NL}
•
###.####
LIM:VOLTage:{HIGH|LOW
}{SP}{NR2}{;|NL}
•
•
•
LIM:VOLTage:{HIGH|LOW
}{?}{;|NL}
•
•
•
###.####
Table 4-6 Measure and Limit Commands
COMMAND
SLM DC
SLD
SLM AC
RETURN
GLOBal:[STATe:] PRESet{SP}{ON|OFF}{;|NL}
•
•
•
GLOBal:[STATe:] LOAD{SP}{ON|OFF}{;|NL}
•
•
•
GLOBal:[STATe:] MODE{SP}{ON|OFF}{;|NL}
•
•
•
GLOBal:[STATe:] SHORt{SP}{ON|OFF}{;|NL}
•
•
GLOBal:[STATe:] DYNamic{SP}{ON|OFF}{;|NL}
•
•
GLOBal:[STATe:] LEVEL{SP}{HIGH|LOW}{;|NL}
•
GLOBal:[STATe:] LEVEL{SP}{A|B}{;|NL}
•
GLOBal:[STATe:] RANGe{SP}{1|2}{;|NL}
•
•
GLOBal:MEASure:CURRent{?}{;|NL}
•
•
•
###.##
GLOBal:MEASure:VOLtage{?}{;|NL}
•
•
•
###.##
Table 4-7 Global Commands