AMETEK SLM Series Rev B User Manual
Page 39
SLM-Series AC/DC Electronic Load
Operation
M540072-01 Rev B
3-13
3.4 INITIAL SETTINGS OF SLM-SERIES AC/DC ELECTRONIC
LOAD
The initial setting parameters of SLM-Series AC/DC Electronic Load are described in
Table 3-2.
Description
SLM-60-20-300 SLM-150-8-300 SLM-300-4-300 SLM-500-1-300
MODE
CC CC CC CC
LOAD OFF
OFF
OFF
OFF
LEVEL LO
LO
LO
LO
SENSE OFF
OFF
OFF
OFF
PRES OFF
OFF
OFF
OFF
WATT OFF
OFF
OFF
OFF
LIMITS
V LIMIT High [V
RMS
] 100.0
200.0
400.0
600.0
ALIMIT High [A
RMS
] 25.00
10.00
5.000
2.000
WLIMIT High [W]
400.0
400.0
400.0
400.0
VALIMIT High [VA]
400.0
400.0
400.0
400.0
All Lower Limits
0.00
0.0
0.0
0.0
C.F
√2
√2
√2
√2
FREQ [Hz]
60
60
60
60
BANK 0
0
0
0
SYNC OFF
OFF
OFF
OFF
CC Level LO [A
RMS
] 0.000
0.000
0.000
0.000
CC Level HI [A
RMS
] 0.000 0.000 0.000 0.000
CR Level LO [
Ω]
4800 30E3 120E3 80E3
CR Level HI [
Ω]
4800 30E3 120E3 80E3
Table 3-3Table 3-2 SLM-Series AC/DC Electronic Load Initial Settings
Last Setting
All SLM-Series AC/DC Electronic Loads retain the last settings used prior to
powering down, so that those same settings are still set at next power-on.
Reset
If the Load’s memory data have been damaged in some way, for example, due to
unstable power source or noises, there may be an error, such as the display
screen showing something different from the actual condition. The Reset
function corrects the errors.
To reset the SLM-Series AC/DC Electronic Load, simultaneously press the
SENSE and the PRES keys. The front panel monitor will display firmware
version and initialize the setting parameters of SLM-Series AC/DC Electronic
Load as shown in Table 3-2 until the key is released.