SLM-Series AC/DC Electronic Load

Operation

M540072-01 Rev B

3-13

3.4 INITIAL SETTINGS OF SLM-SERIES AC/DC ELECTRONIC

LOAD

The initial setting parameters of SLM-Series AC/DC Electronic Load are described in

Table 3-2.

Description

SLM-60-20-300 SLM-150-8-300 SLM-300-4-300 SLM-500-1-300

MODE

CC CC CC CC

LOAD OFF

OFF

OFF

OFF

LEVEL LO

LO

LO

LO

SENSE OFF

OFF

OFF

OFF

PRES OFF

OFF

OFF

OFF

WATT OFF

OFF

OFF

OFF

LIMITS

V LIMIT High [V

RMS

] 100.0

200.0

400.0

600.0

ALIMIT High [A

RMS

] 25.00

10.00

5.000

2.000

WLIMIT High [W]

400.0

400.0

400.0

400.0

VALIMIT High [VA]

400.0

400.0

400.0

400.0

All Lower Limits

0.00

0.0

0.0

0.0

C.F

√2

√2

√2

√2

FREQ [Hz]

60

60

60

60

BANK 0

0

0

0

SYNC OFF

OFF

OFF

OFF

CC Level LO [A

RMS

] 0.000

0.000

0.000

0.000

CC Level HI [A

RMS

] 0.000 0.000 0.000 0.000

CR Level LO [

Ω]

4800 30E3 120E3 80E3

CR Level HI [

Ω]

4800 30E3 120E3 80E3

Table 3-3Table 3-2 SLM-Series AC/DC Electronic Load Initial Settings

Last Setting

All SLM-Series AC/DC Electronic Loads retain the last settings used prior to
powering down, so that those same settings are still set at next power-on.

Reset

If the Load’s memory data have been damaged in some way, for example, due to
unstable power source or noises, there may be an error, such as the display
screen showing something different from the actual condition. The Reset
function corrects the errors.

To reset the SLM-Series AC/DC Electronic Load, simultaneously press the
SENSE and the PRES keys. The front panel monitor will display firmware
version and initialize the setting parameters of SLM-Series AC/DC Electronic
Load as shown in Table 3-2 until the key is released.