Features, Functions, Specifications

SLM-Series AC/DC Load

1-4

M540072-01 Rev B

1.1.2 Operating

Modes

The work mode of SLM-Series AC/DC Electronic Load includes Constant Current (CC)
and Constant Resistance (CR).

CC Mode

During CC mode, the load current input into SLM-Series, AC Electronic Load
depends on the current setting regardless of the input voltage, i.e., the current
setting remains unchanged.

I

V

CC

LOAD

CURRENT

CURRENT SETTING

INPUT VOLTAGE

Figure 1-5 Characteristics of CC Mode

CR Mode

During C.R. mode, the load current input into SLM-Series AC/DC Electronic Load
depends on the resistance setting. At this time, the load current is in direct
proportion to input voltage, e.g. the resistance setting remains unchanged. Please
refer to Figure 1-6 below.

I

V

LOAD

CURRENT

INPUT VOLTAGE

RESISTANCE

SETTING

Figure 1-6 Characteristics of CR Mode

The load setting of SLM-Series AC/DC Electronic Load and the load condition setting of the
front panel can be made through front panel manual operation or through GPIB commands.
The load voltage and current can be transmitted to the computer through GPIB bus. For
operation of GPIB, please refer to Section 4.