Page 24

Installation
SLM-Series AC/DC Load
2-2
M540072-01 Rev B
Figure 2-1 Plug-in Installation and Removal
2.2.2 Removal
1. Turn the SLM mainframe power OFF. Damage may occur to the plug-in
module circuitry if removed under power.
2. Using a screwdriver, loosen the screw on the lower right corner of the front panel.
3. Turn the adjustment knob of the black binding post counter-clockwise until it is
fully loosened.
4. Pull on the black adjustment knob of binding post until the interconnecting jack at
the rear of the load module disengages.
5. Finish pulling the SLM-Series load module out from the mainframe.
2.2.3 Environmental Requirements
• For indoor use only
• Installation Category II (over voltage)
• Pollution Degree 2
• Altitude up to 2000 meters (with power derating)
• Relative Humidity 80% RH Max
Ambient Operating Temperature 0-40
°C, with ideal being 25°C ± 5°C
2.2.4
Service or Repair
If the instrument is damaged, please attach a tag to the instrument, identifying the
owner and indicating the required service. Follow the procedures detailed on Page iii
of this manual.