beautypg.com

AMETEK SLH Series RevB User Manual

Page 59

background image

SLH-Series AC High Power Electronic Load

GPIB/RS232 Programming Operation

M540073-01 Rev C

4-17

WAVEFORM BANK

Purpose:
To read the set value of BANK
Command Syntax:
BANK?{;|NL}
Description:
BANK? Reads back the set value of BANK. 0-10 denotes waveform bank of level 0-10.

WAVEFORM

Purpose:
To read the set value of WAVE.
Command Syntax:
Modules: WAVE?{;|NL}
Description:
WAVE? Reads back the set value of WAVE. 1-5 denotes the C.F. setting of level 1-level 5.

FREQUENCY

Purpose:
To read the set frequency of FREQ.
Command Syntax:
FREQ?{;|NL}
Description:
Reads back the set frequency of FREQ, unit is Hz.

VOLTAGE Limit

Purpose:
To read the set value of upper/lower limit value of threshold voltage.
Command Syntax:
LIM:VOLT:{HIGH/LOW}?{;|NL}
Description:
LIM:VOLT:LOW? Reads back the lower limit set value of threshold voltag;, unit is “volts" (V).

CURRENT Limit

Purpose:
To read the set value of upper/lower limit value of threshold current.
Command Syntax:
LIM:CURR{HIGH|LOW}?{;|NL}
Description:
LIM:CURR:LOW? Reads back the lower limit set value of threshold current; unit is
“amps (A)”.

POWER Limit

Purpose:
To read the set value of upper/lower limit value of threshold power(W).
Command Syntax:
LIM:POW{HIGH|LOW}?{;|NL}
Description: