Page 43

background image

M540073-01 Rev C

4-1

SECTION 4

GPIB/RS-232 PROGRAMMING

OPERATION

4.1 INTRODUCTION

The rear panel of SLH-series is designed for automated and/or remote testing from a
computer through a GPIB or RS-232 interface. This feature can be used as an
automatic load/cross load regulation and centering voltage testing for a switching
power supply or a rechargeable battery charge/discharge characteristic testing.

4.2 SLH-SERIES GPIB COMMAND LIST

4.2.1 Setting

Commands

Function Command Format

SET AUTO FREQUENCY
DETECT ON/OFF

AFREQ{SP}{0|1|OFF|ON}{;|NL}

SET SYNC ON/OFF

SYNC{SP}{0|1|OFF|ON}{;|NL}

SET SENC ON/OFF

SENS{SP}{0|1|OFF|ON}{;|NL}

SET WATT ON/OFF

WATT{SP}{0|1|OFF|ON}{;|NL}

SET BANK

BANK{SP}{d}{;|NL}

d:0 ~ 10

SET WAVE

WAVE{SP}{m}{;|NL}}

m:1 ~ 5

SET FREQUENCY

PREQ{SP}{NR2}{;|NL}

NR2:###.#####

SET VOLTAGE LOW LIMIT

VL{SP}{NR2}{;|NL}

SET VOLTAGE HIGH LIMIT

VH{SP}{NR2}{;|NL}

SET CURRENT LOW LIMIT

IL{SP}{NR2}{;|NL}

SET CURRENT HIGH LIMIT

IH{SP}{NR2}{;|NL}

SET WATT LOW LIMIT

WL{SP}{NR2}{;|NL}

SET WATT HIGH LIMIT

WH{SP}{NR2}{;|NL}

SET VA LOW LIMIT

VAL{SP}{NR2}{;|NL}

SET VA HIGH LIMIT

VAH{SP}{NR2}{;|NL}

SET CURRENT

CURR:{LOW|HIGH|A|B}{SP}{NR2}{;|NL} NR2:###.#####

SET RESISTANCE

RES:{LOW|HIGH|A|B}{SP}{NR2}{;|NL}

SET LOAD ON/OFF

LOAD{SP}{0|1|OFF|ON}{;|NL}

SET LEVEL LOW/HIGH

LEV{SP}{0|1|LOW|HIGH}{;|NL}

SET PRESENT ON/OFF

PRES{SP}{0|1|OFF|ON}{;|NL}

SET MODE

MODE{SP}{0|1|2|CC|CR|CV}{;|NL}

SET CLEAR STATE

CLER{;|NL}

SET STORE STATE

STORE{SP}{m}{;|NL}

m:1 ~ 5

SET RECALL STATE

RECALL{SP}{m}{;|NL}

m:1 ~ 5

Table 4-1 GPIB Setting Command Summary