SLH-Series AC High Power Electronic Load

GPIB/RS232 Programming Operation

M540073-01 Rev C

4-7

LOAD

Purpose:
Turn the Electronic load module input ON or OFF.
Command Syntax:
LOAD{SP}{0FF|ON}{NL}
Description:
This command sets the Electronic load to sink current from power source.
For Example :
LOAD OFF ; All loads are at input OFF condition.
LOAD ON ; Set the load to LOAD ON status, this load is ready to sink current from the
power source.
LOAD 0; Set the load to LOAD OFF.

LEVEL HIGH/LOW

Purpose:
Select Low or High level in static mode, of DC electronic loads, or LEVEL A/B of AC
electronic loads.
Command Syntax:
[GLOB:] LEVE {SP}{HIGH|LOW}{NL}
Description:
LEVE LOW is Set LOW current level in CC mode or LOW resistance level in CR mode.
LEVE 1 is Set HIGH current level in CC mode or HIGH resistance level in CR mode.

PRESET ON/OFF

Purpose:
Set the upper or lower digit multi-function meter to display the programming load level.
Command Syntax:
PRES{SP}{0|1|OFF|ON}{NL}
Description:
PRES ON is set to preset on status.

MODE

Purpose:
Select the operating mode of Electronic load module.
Command Syntax:
MODE{SP}{0|1|CC|CR}{NL}
Description:
MODE CC ; set the presently operating mode to Constant Current mode.
MODE 1 ; set the presently operating mode to Constant Resistance mode.

CLEAR status register

Purpose:
CLEar the PROT and ERR status byte registers.
Command Syntax:
All Modules:

CLER{NL}