REPEAT

This is the number of times the SURGE/SAG transient event
will repeat before it will proceed to the next event or exit the
transient program. Note that the number of times the transient
event is generated is equal to the REPEAT + 1. Leave this
value at zero if only one execution of this event in the list is
required.

EVENT #

This must be the last item in the transient edit screen. All data
fields must be entered before inserting the EVENT #. The
EVENT # takes a value from 1 to 99. The EVENT # defines the
order of execution of the transient events in a multiple event
transient. It is a good practice to enter spaced EVENT #’s to
allow insertion of an EVENT later if needed. (For example,
space them by 5.) Entry of a sequence EVENT # number will
cause the display to return to the TRANSIENT screen.