User Manual

AMETEK Programmable Power

RS Series

153

6.4.5 Output Calibration Summary

The following Table is a summary of the preceding calibration steps. Program the following
values in the table and make the adjustments in the OUTPUT CALIBRATION screen. Select the
phase to be calibrated by pressing the PHASE key.

CALIBRATION

PROGRAM VALUES

CALIBRATION VALUE

ADJUST TO

300 VAC range DC Zero

300 VAC range, 0.0 V,
60 Hz

VOLT ZERO

0 ± 5 mV DC

300 VAC range High Freq
DC Zero, -HF models only.

300 VAC range, 0.0 V,
820 Hz

VOLT ZERO

0 ± 5 mV DC

300 VAC range Volt FS

240.0 V, 60 Hz

VOLT FS

240

±

0.05 VAC

400 VDC range DC Zero

400 VDC range, 0.0 V

VOLT ZERO

0 ± 5 mV DC

Phase Offset B, C

150 VAC range, 120V,
60 Hz

PHASE OFST

± 0.5

°

Option -413

400 Hz, 20VAC
1800 Hz, 20VAC

R9, R10, R11
INTER HARM FS

20 ± 1.0 Vrms
20 ± 1.0 Vrms

Table 6-3: Output Calibration Table – RS Series