9.3 Option –160: RTCA/DO-160 Tests

The RTCA/DO-160 Option is made up of both firmware that resides in the power source and the
RSGui Windows application program. The firmware covers revision D and can be used from the
front panel or under RSGui control. Revision E is implemented through the RSGui only.

The user interface for each implementation is different. The revision D tests can be operated
directly from the power source’s front panel or through the supplied RSGui program. The
Revision E tests can only be operated through the RSGui windows software. Thus, for Rev E
use, a Windows PC and interface – RS232 ,USB, LAN or GPIB (recommend) - is required.

Section 9.3.1 covers operation of the firmware based revision D tests.

For information regarding the operation of the DO160 revision E tests with the RSGui, please
refer to the Avionics Software Manual (CI part no. 4994-971 included on CDROM). Note that
future updates of the RSGui may include overlapping coverage for revision D in the software as
the RSGui program is designed to support all revisions. For now, revision D has no associated
data files in the RSGui but does have it’s own user interface control screen.

9.3.1 Option –160: RTCA/DO160 rev D Tests (Firmware)

9.3.1.1 General

The RTCA/DO-160 option is capable of performing all sections of RTCA/DO-160D, Section 16
and EUROCAE-14D / RTCA DO160D, Section 16 for the AC Source signal. A selection is made
available to specify the type of standard, and groups.

Through out this document, Groups 1 through 3 will be used to refer to EUROCAE-14D
standard.

9.3.1.2 Initial Setup

Nominal parameters for the AC Power source are as follows:

Output Voltage

115V L-N or 230V L-N

Output Frequency

360 Hz to 800 Hz

Nominal parameters for the DC Power source are as follows:

Output Voltage

28V or 14V L-N

A setting outside these nominal values will disable the test and will prevent accessing to its
Menu screens. To access the test for the 230V L-N, the power source must be capable of
programming 360V.

9.3.1.3 Tests Performed

Available tests are divided into Normal, Abnormal and Emergency.

9.3.1.3.1 NORMAL STATE

AC Source:

1. Normal State Voltage and Frequency test

2. Voltage unbalance test
3. Waveform Distortion test

4. Voltage Modulation test

5. Frequency Modulation test

6. Momentary Power Interrupt (Undervoltage) test