4 limit values, Limit values – Pilz PSSu E S 2AI U User Manual
Page 19

Function description
Operating Manual PSSu E S 2AI U(T)
21405EN03
19
4.2.4
Limit values
The module has range monitoring and limit value monitoring on each channel:
}
Range monitoring
}
Upper limit value: 4095
D
–
Lower limit value: 4095
D
–
The module compares the upper and lower limit value with the digital value after the
hardware calibration (values with 12 bits plus sign) and writes the result of the com
parison as follows:
– System environment A:
– In the status byte (see "PSSu assignment in system environment A")
– System environment B:
– In the I/O data element "Overrange" or "Underrange" (see "PSSu assignment in
system environment B").
–
The limit of the measuring range corresponds to 4095
D
.
}
Limit value monitoring
}
Limit value 1
–
Limit value 2
–
The module compares limit value 1 and limit value 2 with the digital value after scal
ing (values with 15 bits plus sign) and writes the result of the comparison as follows:
– System environment A:
– In the status byte (see "PSSu assignment in system environment A")
– System environment B:
– In the I/O data element "LimitValue1" or "LimitValue2" (see "PSSu assignment in
system environment B").
–
The limit of the measuring range corresponds to 32 767
D
with default scaling val
ues.
You can change the default values in the system software.
The decimal value for the PSSuniversal Configurator (n) is calculated from the analogue
value at the input (U
Limit
) as follows:
n = 32 768 * U
Limit
/ 10 V
Example:
}
Voltages at the input, which are to be monitored through the limit values:
–
Limit value 1 is to be 8 V.
–
Limit value 2 is to be 5 V.
}
Entry in the PSSuniversal Configurator:
–
Limit value 1 corresponds to 26 214
–
Limit value 2 corresponds to 16 384