Acid is used as a test sample, Standardization – Parr Instrument 6100 User Manual

Page 57

background image

Standardization

6100

C

w w w . p a r r i n s t . c o m

55

Table C-3

Calorimeter Control Limit Values in BTU/lb When Benzoic Acid is Used as a Test Sample

Accepted heat of combustion taken as 11373 BTU/lb.
Instrument precision 0.20%.
Control limits based on 99% confidence (3 sigma) values.
Values are in BTU/lb.

NUMBER OF

OBSERVATIONS

IN A GROUP

UCL FOR THE RANGE

(HIGH – LOW) WITHIN

THE GROUP

UCL FOR THE

RSD WITHIN THE

GROUP

MAXIMUM PERMISSIBLE DEVIATION

OF THE GROUP MEAN FROM THE

ACCEPTED VALUE OR GRAND MEAN

1

68.2

2

83.9

0.521%

48.3

3

99.1

0.455%

39.4

4

106.9

0.418%

34.1

5

111.9

0.393%

30.5

6

115.5

0.375%

27.9

7

118.3

0.361%

25.8

8

120.7

0.350%

24.1

9

122.7

0.341%

22.7

10

124.4

0.334%

21.6

11

125.9

0.327%

20.6

12

127.2

0.322%

19.7

13

128.4

0.317%

18.9

14

129.6

0.313%

18.2

15

130.6

0.309%

17.6

16

131.5

0.305%

17.1

17

132.4

0.302%

16.6

18

133.2

0.299%

16.1

19

133.9

0.297%

15.7

20

134.7

0.294%

15.3

21

135.3

0.292%

14.9

22

136.0

0.290%

14.5

23

136.6

0.288%

14.2

24

137.2

0.286%

13.9

25

137.7

0.284%

13.6