Factory default settings, Program installation & control – Parr Instrument 6100 User Manual
Page 18

4
Program Installation & Control
P a r r I n s t r u m e n t C o m p a n y
16
Calorimeter Operations
Operating Mode
Determination
Bomb Installed/EE
1/2400.0
Operating Controls
Method of Operation
Dynamic
Reporting Units
BTU/lb
Use Spiking Correction
OFF
“OTHER” Multiplier
4.1868
Calibrate Touchscreen
LCD Backlight Timeout(s)
1200 s
LCD Backlight Intensity
70%
Print Error Messages
ON
Language
English
Spike Controls
Use Spiking
OFF
Heat of Combustion of Spike
6318.4
Use Fixed Spike
OFF
Weight of Fixed Spike
0.0
Prompt for Spike before Weight
OFF
Program Information and Controls
Date & Time Settings
Software and Hardware Info
Settings Protect
OFF
User/Factory Settings
Feature Key
Bomb Type Select
User Function Setup
Cold Restart
User/Factory Settings
User Setup ID
61-1108
Reload Factory Default Settings
Reload User Default Settings
Save User Default Settings
Calibration Data & Controls
Calibration Run Limit
10
EE Max Std Deviation
0.0
Heat of Combustion of Standard
6318.4
Bomb Service Interval
500
Control Chart Parameters
Use Bomb
1
Bomb 1 Through 4
EE Value
2400.0
Protected EE Value
OFF
Thermochemical Corrections Standardization
Fixed Fuse Correction
ON 50
Acid Correction
Fixed HNO
3
10.0
Fixed Sulfur Correction
ON 0.0
Heat of Formation Sulfuric Acid
36.1
Heat of Formation Nitric Acid
14.1
Determination
Fixed Fuse Correction
ON 50
Acid Correction
Fixed HNO
3
10.0
Fixed Sulfur Correction
OFF 0.0
Calculation Factors
Nitric Acid Factor
1.58
Acid Multiplier
0.0709
Sulfur Value is Percent
ON
Sulfur Multiplier
0.6238
Fuse Multiplier
1.0
Use Offset Correction (ISO)
OFF
Offset Value
0.0
Table 4-1
Factory Default Settings