Menu Operating Instructions

P a r r I n s t r u m e n t C o m p a n y

44

File Management

Run Data File Manager: This key activates the File
Manager. The File Manager is used to delete or
rename test report files. It is also used to convert file
types.

Format the SD Card: This key allows the user to
format an installed SD card in a manner that is
compatible with the calorimeter.

Note:

Formatting will erase all files on the

card!

Copy Run Data to SD Card: This key copies all test
data to an SD card inserted into the rear of the calo-
rimeter controller. This feature is used as a means
of either archiving data or transferring it to a PC.

Note:

Subsequent use of the same SD card

will overwrite the data currently on the card.

Copy User Settings to SD Card: This key copies all
previously saved user setups to SD.

Copy User Settings From SD Card: This key copies
all user setups previously saved to SD back to the
calorimeter controller memory. This feature can be
used to configure multiple calorimeters in an identi-
cal manner.

Diagnostics Menu

Provides the user with the means to test many of the
components and subsystems of the calorimeter.

Test Ignition Circuit: The key activates the ignition
circuit. A volt meter can be placed across the ignition
leads to ensure that the actual firing charge is reach-
ing these contacts.

Data Logger: This key displays and leads to sub-
menus which control the data logging function of the
calorimeter.

View System Log: This key is used to display the con-
tents of /flash/log/messages. This file is used primarily
to log application program debug messages.

User Defined Functions: This key leads to a sub-
menu that offers five special purpose user/factory
definable function keys.

Instrument Monitor: This screen provides a summary
of important instrument parameters. The monitor is
used to detail the course of a test or to observe the
heating/cooling performance of the calorimeter.

View System Info: This key accesses current program
information and settings such as: Processes and their
associated PIDs (proportional (P), the integral (I), and
the derivative (D) controls), memory, mass storage,
network.

View Instrument Log: This screen displays the
contents of tmp/instlog. This file, among other things,
is the logfile destination for the data logger.

I/O Diagnostics: This key accesses a sub-menu which
allows the user to manipulate digital outputs for
troubleshooting.