beautypg.com

Program installation & control – Parr Instrument 6100 User Manual

Page 19

background image

Program Installation & Control

6100

4

w w w . p a r r i n s t . c o m

17

Net Heat/Dry Factors

Fixed Hydrogen

OFF 0.0

Fixed Oxygen

ON 0.0

Fixed Nitrogen

ON 0.0

Calculate Net Heat of Combustion

OFF

Fixed Moisture as Determined

OFF 0.0

Fixed Moisture as Received

OFF 0.0

Dry Calculation

OFF

Data Entry Controls

Prompt for Bomb ID

ON

Weight Entry Mode

Touch Screen

Acid Entry Mode

Touch Screen

Net Heat Entry Values

Touch Screen

Auto Sample ID Controls

ON

Sample Weight Warning above

2.0

Spike Weight Entry Mode

Touch Screen

Sulfur Entry Mode

Touch Screen

Moisture Entry Modes

Touch Screen

Auto Preweigh Controls

ON

Auto Sample ID Controls

Automatic Sample ID

ON

Automatic Sample ID Increment

1

Automatic Sample ID Number

1

Auto Preweigh Controls

Automatic Preweigh ID

ON

Automatic Preweigh ID Increment

1

Automatic Preweigh ID Number

1

Reporting Controls

Report Width

40

Automatic Reporting

ON

Auto Report Destination

Printer

Individual Printed Reports

OFF

Edit Final Reports

OFF

Recalculate Final Reports

OFF

Use New EE Values in Recalculation

OFF

Communication Controls

Printer Type

Parr 1758

Balance Port
Network Interface
Printer Destination

Local (USB)

Bar Code Port
Network Data Devices

Balance Port Communications

Balance Type

Generic

Balance Port Device

/dev/ttyUSB0

Customize Balance Settings

Balance Port Settings

Number of Data Bits

8

Parity

None

Number of Stop Bits

1

Handshaking

None

Baud Rate

9600

Data Characters from Balance

8

Data Precision

4

Transfer Timeout (seconds)

10

Balance Handler Strings

Data Logger

Data Logger

OFF

Data Log Interval

10s

Data Log Destination Log File and Printer
Select Data Log Items
Data Log Format

Text Format