Acid is used as a test sample, Standardization – Parr Instrument 6100 User Manual

Page 56

background image

C

Standardization

P a r r I n s t r u m e n t C o m p a n y

54

Table C-2

Calorimeter Control Limit Values in cal/g When Benzoic Acid is Used as a Test Sample

Accepted heat of combustion taken as 6318 cal/g.
Instrument precision 0.20%.
Control limits based on 99% confidence (3 sigma) values.
Values are in cal/g.

NUMBER OF

OBSERVATIONS

IN A GROUP

UCL FOR THE RANGE

(HIGH – LOW) WITHIN

THE GROUP

UCL FOR THE

RSD WITHIN THE

GROUP

MAXIMUM PERMISSIBLE DEVIATION

OF THE GROUP MEAN FROM THE

ACCEPTED VALUE OR GRAND MEAN

1

37.9

2

46.6

0.521%

26.8

3

55.1

0.455%

21.9

4

59.4

0.418%

19.0

5

62.1

0.393%

17.0

6

64.2

0.375%

15.5

7

65.7

0.361%

14.3

8

67.1

0.350%

13.4

9

68.2

0.341%

12.6

10

69.1

0.334%

12.0

11

69.9

0.327%

11.4

12

70.7

0.322%

10.9

13

71.3

0.317%

10.5

14

72.0

0.313%

10.1

15

72.5

0.309%

9.8

16

73.1

0.305%

9.5

17

73.5

0.302%

9.2

18

74.0

0.299%

8.9

19

74.4

0.297%

8.7

20

74.8

0.294%

8.5

21

75.2

0.292%

8.3

22

75.6

0.290%

8.1

23

75.9

0.288%

7.9

24

76.2

0.286%

7.7

25

76.5

0.284%

7.6