Instrument Description
6100
3
w w w . p a r r i n s t . c o m
13
Menu Keys
The controls that change the data field information in the menus will be one of the following:
1. Toggles: These data fields contain ON/OFF or YES/NO choices. Simply touching the key
on the screen toggles the choice to the other option. The current setting is displayed in the
lower right corner of the key.
2. Option Selection: These data fields contain a list of options. Touching the key on the
screen steps the user through the available choices. The current setting is displayed in the
lower right corner of the key.
3. Value Entry Fields: These data fields are used to enter data into the calorimeter. Touching
the key on the screen brings up a sub menu with a key pad or similar screen for entering
the required value. Some keys lead to multiple choices. Always clear the current value be-
fore entering a new value. Once entered the screen will revert to the previous menu and
the new value will be displayed in the lower right corner of the key.
4. Data Displays: Most of these keys display values that have been calculated by the calorim-
eter and are informational only. Certain ones can be overridden by the user entering a de-
sired value through a sub-menu. The value is displayed in the lower right corner of the key.
Note:
Some keys will respond with an opportunity for the user to confirm the speci-
fied action to minimize accidental disruptions to the program and/or stored data.
Control Keys
There are five control keys which always appear in the right column of the primary displays.
These keys are unavailable when they are gray instead of white.
1. Escape: This key is used to go up one level in the menu structure.
2. Main Menu: This key is used to return to the main menu touch screen from anywhere in
the menu structure.
3. Start: This key is used to start a calorimeter test.
4. Report: This key is used to access the test results stored in the calorimeter, to enter ther-
mochemical corrections and to initiate a report on the display or printer.
5. Help: This key is used to access help screens related to the menu currently displayed on
the touch screen.
6. Abort: This key appears in the start key location while the test is running. Pressing this
key will abort the test in progress.
7.
This key appears on the main menu only and is used to prepare the calorimeter for
turning off the power.