Perating, Nstructions – Parr Instrument 6300 User Manual
Page 70

A
M
ENU
O
PERATING
I
NSTRUCTIONS
P a r r I n s t r u m e n t C o m p a n y
70
D
IAGNOSTICS
M
ENU
(C
ONTINUED
)
Combine Det. Reports.
•
Pressing this key com-
bines all determination reports into a single file
named /tmp/bigdetfile.txt.
Combine Std. Reports.
•
Pressing this key com-
bines all determination reports into a single file
named /tmp/bigstdfile.txt.
Rinse Bomb:
This key initiates a bomb rinse. This function can be
used to clean out the cylinder in the event a sample is
spilled inside the cylinder.
Instrument Monitor:
This key accesses its sub-menu screen which provides a
summary of most of the important instrument param-
eters. This screen is used to detail the course of a test
or to observe the heating/cooling performance of the
calorimeter.
View System Info:
Press this key to display a screen with current operating
system information/statistics such as:
Processes and their associated PIDs
•
Memory
•
Mass Storage
•
Network
•
Press the PRINT key to print this information.
View Instrument Log:
Press this key to display a screen with contents of the
/tmp/instlog file. It contains a sequential log of the
instrument’s processing. Press the PRINT key to print
this log.
I/O Diagnostics:
Press this key to display the I/O Diagnostics sub-menu,
which allows the user to manipulate digital outputs for
troubleshooting. The I/O Diagnostics screen is used
to display the digital outputs at a basic level for trouble-
shooting. Both the bucket and jacket temperatures are
also displayed on this screen. Any output can be
selected using the left and right arrow keys. The select-
ed output is turned ON (1) or OFF (0) using the 1 and 0
keys. Prior to entering the Diagnostics Menu, the con-
troller stores the present state of the outputs. This state
is restored when you exit this screen. Digital outputs
cannot be manipulated while a test is in progress.