The 6300 Calorimeter will hold data for 1000 tests in its
memory. These tests may be pre weights, preliminary or
final reports for either standardization or determination
runs. Once the memory of the controller is filled, the
controller will not start a new analysis until the user
clears some of the memory.

C

LEARING

M

EMORY

The FILE MANAGEMENT key on the main menu
leads to the file management sub-menu. The RUN
DATA FILE MANAGER key leads to a listing of the
files.

Single files can be deleted by highlighting the file

and pressing the DELETE key. The controller will
then ask the user to confirm that this file is to be
deleted.
A series of files can be deleted by selecting the first

file in the series and then the last file in the series
using the EXTEND SEL key and then pressing the
DELETE key.

R

EMOVABLE

SD M

EMORY

The controller of the 6300 calorimeter can accept SD
memory cards. These cards can be used to:

Copy test file data for transfer to a computer.

Copy user settings for back up.

Reload user settings to the controller to restore or

update the controller’s operating system.

SD memory cards are inserted into the slots on the
back of the control section of the calorimeter. Keys
are provided on the FILE MANAGEMENT sub-
menu to initiate each of the above three actions with
the exception of restoring or updating the controllers
operating system.