beautypg.com

Figure g-1, Vessel rinse container — 128, Inse – Parr Instrument 6300 User Manual

Page 128: Ontainer, Ssembly

background image

G

B

OMB

R

INSE

C

ONTAINER

A

SSEMBLY

P a r r I n s t r u m e n t C o m p a n y

128

O

PERATION

(C

ONTINUED

)

Samples to be analyzed for chlorine should be spiked in
a manner to insure that the sample as-burned contains
less than 10% chlorine and liberates at least 5000 calo-
ries. Five thousand calories corresponds to roughly a
6.4 °C temperature rise in a 6300 Calorimeter.

Figure G-1

Vessel Rinse Container

1051DD

EXHAUST DIFFUSER

308VBAD MALE CONNECTOR

(1/4T-1/4NPTM)

HX0025TB062

1/4" TUBING X 4 FT LG

411VBAD REDUCER

(1/4T-3/8Tx)

221VBAD 3/8T NUT

222VBAD 3/8T FERRULE SET

1052DD

BOTTLE CAP

A138CA MALE CONNECTOR

(1/4T-1/8NPTM)

1.50

1053DD BOTTLE, 500ml

At the completion of the test, the collection bottle
should be unscrewed from the 1052DD Bottle Cap and
the contents of the combustion capsule added to the
collected washings. The bottle then may be capped and
labeled until the contents can be analyzed.

Most users find it useful to add the contents of the
rinsed combustion capsule to the washings collected in
the bottle.