Data Precision. This key allows the user to
»
establish the number of digits to the right of
the decimal point that are passed from the
balance handler.
Transfer Timeout (seconds). This value
»
determines how long the interface will wait
before giving up on a weight transfer. The
value is entered in seconds.
Balance Handler Strings. This key leads to a
»
submenu that allows balance template to be
customized for unique balances or needs.
Log Balance to Display.
•
Directs the incoming
data stream from the balance to a display buf-
fer. This function can be used to determine the
data format from an unknown balance type.
The display buffer is 40 characters in length.
The balance must be forced to issue at least 40
characters before the contents of the buffer are
displayed.
Balance Port LoopBack Test.
•
Initiates a loopback
test on the port. A special loopback plug is
required in order to perform this test.
Further information on establishing communications
for the Printer, Balance, Network Interface, Bar
Code and other Network Data Devices can be found
in Appendix D, Communication Interfaces, of this
manual.
F
ILE
M
ANAGEMENT
Run Data File Manager:
This key activates the File Manager. The File Manager
is used to delete or rename test report files. It is also
used to convert file types.
Format the SD Card:
This key allows the user to format an installed SD card
in a manner that is compatible with the calorimeter.
Note:
Formatting will erase all files on the card!
Copy Run Data to SD Card:
This key copies all test data to an SD card inserted into
the rear of the calorimeter controller. This feature is
used as a means of either archiving data or transferring
it to a PC.
Note:
Subsequent use of the same SD card will overwrite the
data currently on the card.
Copy User Settings to SD Card:
This key copies all previously saved user setups to SD.
Copy User Settings From SD Card:
This key copies all user setups previously saved to SD
back to the calorimeter controller memory. This feature
can be used to configure multiple calorimeters in an
identical manner.