The program in the 6300 Calorimeter can be extensively
modified to tailor the unit to a wide variety of operating
conditions, reporting units, laboratory techniques,
available accessories and communication modes.

In addition, the calculations, thermochemical
corrections and reporting modes can be modified to
conform to a number of standard test methods and
procedures.

Numerous provisions are included to permit the use of
other reagent concentrations, techniques, combustion
aids and short cuts appropriate for the user’s work.

Note:

Changes to the program are made by use of the menu
structure described in Appendix A of this manual. Any
of these items can be individually entered at any time to
revise the operating program.

D

EFAULT

S

ETTINGS

Units are pre programmed with DEFAULT
SETTINGS. See pages 24 and 25 for a listing of the
factory default settings.

These default settings remain in effect until changed
by the user. Should the user ever wish to return to the
factory default settings, go to the Program Information
and Control Menu, then to User/Factory Settings, and
then touch Reload Factory Default Settings and YES.

Non-volatile memory is provided to retain any and
all operator initiated program changes; even if power
is interrupted or the unit is turned off. If the unit
experiences an intentional or unintentional “Cold
Restart”, the controller will return to its default settings.

R

EVISING

D

EFAULT

S

ETTINGS

The default parameters of the 6300 Calorimeter can
be changed to guarantee that the 6300 Calorimeter,
when cold restarted, will always be in the desired
configuration before beginning a series of tests.

Users who wish to permanently revise their default
settings may do so using the following procedure:

Establish the operating parameters to be stored as

the user default settings.

Go to the Program Info and Control Menu, User/

Factory Settings, User Setup ID, and enter the
desired User Setup ID.

Select Save User Default Settings

To re-load the user default settings, go to the Program
Info and Control Page, User/Factory Settings, Re-load
User Default Settings, and YES.