8 Maintenance and service

60

Version 11/2006, Rev. 2.0 of 02/12/2014 • sb

Translation of the original operating manual

8.1.5 Accessories

CAUTION

For the care of the accessories, special safety measures must be
considered as these are measures that will ensure operational safety at
the same time.

Chemical reactions as well as stress-corrosion (combination of oscillating
pressure and chemical reaction) can affect or destroy the metal and plastic
parts. Barely detectable cracks on the surface can expand and weaken the
material without any visible signs.
• Check the material regularly (at least once a month) for

– cracks
– visible damage of the surface
– pressure marks
– signs of corrosion
– other changes.

• Replace any damaged components immediately for your own safety.
• Immediately rinse off the accessories if any liquids that may cause

corrosion come into contact with them.

• Clean the accessories outside the freeze-dryer once a week or

preferably after each use.

Pos: 184 /01 0 Univ ersal mod ule/Seit enwec hsel @ 0\m od_ 120 211 624 4312 _0. docx @ 10 5 @ @ 1

Pos: 185 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 80 Wa rtun g u nd In stan dhalt ung/ 080 -00 20 Desinf ektion von Tr ockn ungsk am mer und Zu beh ör @ 25\ mo d_1 4049 841 078 51_ 68.d ocx @ 184 461 @ 2 @ 1

8.2

Disinfection of the drying chamber and accessories

• Use commercially-available disinfectants such as, for example, Incidur ,

Meliseptol , Sagrotan , Buraton , or Terralin (available at specialised
trade).

• The freeze-dryers and the accessories consist of various materials. A

possible incompatibility must be considered.

• Before using cleaning or decontamination agents that were not

recommended by us, contact the manufacturer to ensure that such a
procedure will not damage the freeze-dryer.

• Please contact us if you have any queries (see chapter 7.3 - "Service

contact").

DANGER

If dangerous materials (e.g. infectious and pathogenic substances) are
used, the freeze-dryer and accessories must be disinfected.