8 Maintenance and service
Version 11/2006, Rev. 2.0 of 02/12/2014 • sb
57
Translation of the original operating manual
Pos: 169 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 80 Wa rtun g u nd In stan dhalt ung/ 080 -00 10- 001 0-0 010 Eiskon dens ato rkam me r (a uß er Epsil on 2 -4) @ 2 5\m od_ 1404 984 098 898 _68. docx @ 18 434 9 @ 4 @ 1
8.1.1.1 Ice condenser chamber
Before each start-up, ensure that the ice condenser chamber is free from
water residues.
• Open the media drain valve to drain off any liquid. Then, close the
valve.
• If necessary, wipe the ice condenser chamber dry with a cloth.
Pos: 170 /01 0 Univ ersal mod ule/L eerz eile @ 0\m od_ 120 211 624 450 0_0. docx @ 11 4 @ @ 1
Pos: 171 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 80 Wa rtun g u nd In stan dhalt ung/ 080 -00 10- 001 0-0 020 T rock nun gska mme r La bor @ 2 5\m od_ 140 4984 100 052 _68. docx @ 18 436 3 @ 4 @ 1
8.1.1.2 Drying chamber
Some drying chambers are equipped with the LYO CHAMBER GUARD
splinter protection film.
• In order to prevent the film from being damaged, only use water and
commercially available window cleaning agents as well as synthetic
sponges, soft tissues or a rubber wiper for cleaning.
• Do not dry-clean any drying chambers that are equipped with LYO
CHAMBER GUARD film!
Fig. 45: Identification label on a drying chamber with a splinter protection film
Pos: 172 /01 0 Univ ersal mod ule/Seit enwec hsel @ 0\m od_ 120 211 624 4312 _0. docx @ 10 5 @ @ 1
Pos: 173 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 80 Wa rtun g u nd In stan dhalt ung/ 080 -00 10- 001 0-0 030 Belüf tung s- u nd Medi ena blaufv entil LD plus @ 31\ mo d_14 084 423 160 92_ 68.d ocx @ 209 746 @ 4 @ 1
8.1.1.3 Aeration valve, media drain valve
Contaminants such as product residues may lead to an insufficient vacuum.
In this case, the aeration valve and the media drain valve must be cleaned.
• Switch the freeze-dryer off and disconnect the mains plug.
• Remove the valve core.
• Clean the valve core and the opening with a moist cloth.
• Clean the O-rings and inspect them for any damage. Damaged O-rings
must be replaced.
1
Valve opening
2
Valve core
3
O-rings
Fig. 46: Valve opening and valve core with O-rings
(example, varies depending of the type of freeze-dryer)