5 set-up and connection, 1 installation site, 2 power supply – Martin Christ Alpha 1-2 LDplus User Manual
Page 32

Freeze-dryer Alpha 1-2 LDplus
5 Set-up and connection
32
Version 11/2006, Rev. 2.0 of 02/12/2014 • sb
Translation of the original operating manual
Pos: 92 / 200 Christ /36 0 GT -BA La bor -Pilot (S TANDARDMODULE) /05 0 Aufstell ung und Anschlus s/05 0 Aufs tellun g u nd Ansc hluss= == == === == == === == == == === == @ 25 \mo d_1 404 983 7372 44_ 68. docx @ 18 3495 @ 1 @ 1
5 Set-up and connection
Pos: 93 / 200 Christ /36 0 GT -BA La bor -Pilot (S TANDARDMODULE) /05 0 Aufstell ung und Anschlus s/05 0-0 010 Aufst ellort (Alpha -Delt a) @ 25\ mod _14 049 837 381 97_ 68.d ocx @ 183 509 @ 2 @ 1
5.1 Installation site
Operate the freeze-dryer only in closed and dry rooms.
• The table must be stable and have a solid, even tabletop.
• Ensure sufficient ventilation. Do not place any paper, cloth, or similar
material behind or under the unit, since otherwise the air circulation will
be impaired.
• Keep a safety distance of at least 30 cm from the wall so that the vents
in the unit remain fully effective.
• The ambient temperature must be in the range of +10°C to +25°C.
• Do not subject the freeze-dryer to thermal stress, e.g. by positioning it
near heat generators.
• Avoid direct sunlight (UV radiation).
Pos: 94 / 010 Unive rsalm odul e/Le erzeil e @ 0\mo d_1 202 116 244 500 _0.d ocx @ 114 @ @ 1
Pos: 95 / 200 Christ /36 0 GT -BA La bor -Pilot (S TANDARDMODULE) /05 0 Aufstell ung und Anschlus s/05 0-0 020 Ene rgiev erso rgu ng- -- --- --- -- --- -- --- --- -- --- -- --- -- --- -- @ 25\ mo d_14 049 837 400 08_ 68.d ocx @ 183 537 @ 2 @ 1
5.2 Power supply
Pos: 96 / 200 Christ /36 1 GT -BA La bor -Pilot (PROJEKTE )/Alpha 1- 2 LDpl us/0 50 Aufs tellun g un d Anschl uss/0 50- 002 0-0 010 Anschl ussar t Alpha 1- 2 LDpl us @ 31\ mod _140 808 408 036 5_6 8.doc x @ 2 093 98 @ 3 @ 1
5.2.1 Connection
DANGER
The operating voltage on the name plate must correspond to the local
supply voltage!
CHRIST freeze-dryers are units of safety class I.
Alpha 1-2 LDplus units have a three-wire power cord with an IEC C13
connector (see chapter 10 - "Technical data").
An equipotential bonding screw is located on the back below the mains
power input (see chapter 2.1.1 - "Functional and operating elements"). This
equipotential bonding screw can be used to perform an earth conductor
check.
Pos: 97 / 010 Unive rsalm odul e/Le erzeil e @ 0\mo d_1 202 116 244 500 _0.d ocx @ 114 @ @ 1
Pos: 98 / 200 Christ /36 0 GT -BA La bor -Pilot (S TANDARDMODULE) /05 0 Aufstell ung und Anschlus s/05 0-0 020 -00 20 Sic her unge n b auseits @ 2 5\m od_ 140 4983 741 015 _68. docx @ 18 355 1 @ 3 @ 1
5.2.2 Customer-provided fuses
Typically, the freeze-dryer must be protected with 16 Amp G fuses that are
to be provided by the customer.
Pos: 99 / 010 Unive rsalm odul e/Seite nwechs el @ 0 \mo d_1 202 116 244 312_ 0.d ocx @ 105 @ @ 1