beautypg.com

3 service contact – Martin Christ Alpha 1-2 LDplus User Manual

Page 55

background image

Freeze-dryer Alpha 1-2 LDplus

7 Malfunctions and error correction

Version 11/2006, Rev. 2.0 of 02/12/2014 • sb

55

Translation of the original operating manual

Error message

Possible cause

Remedy

Ice condenser temperature 20°C
The ice condenser temperature rises to
a non-permissible value during the
freeze-drying process. The pressure
control valve will close automatically in
order to prevent the vacuum pump from

being damaged.

• Insufficient cooling.

• Observe the behaviour of the

refrigeration system. If
necessary, inform the service
department (see chapter 7.3 -

"Service contact").

Defective ice condenser temperature

sensor

• The temperature sensor is not

connected or it is defective.

• Inform the service department

(see chapter 7.3 - "Service
contact").

Defective temperature sensor of the

defrosting heater

• The temperature sensor is not

connected or it is defective.

• Inform the service department

(see chapter 7.3 - "Service
contact").

Mains power failure (see chapter 7.1.1

- "Power failure")

• The power supply has been

interrupted.

• The mains power connector is

not connected.

• The fuses have tripped.

• The mains power switch has

been switched off.

• Check the mains fuse.

• Connect the mains power

connector firmly.

• Check the customer-provided

fuses.

• Switch the mains power switch

on.

Default settings loaded
All of the settings have been reset to the

delivery state of the freeze-dryer.

• Error in the parameter memory.
• Manual execution in the service

menu.

• No measure required.

IO communication error (0x20)

• Error in the LDplus IO module.

• Inform the service department

(see chapter 7.3 - "Service
contact").

Pos: 159 /01 0 Univ ersal mod ule/L eerz eile @ 0\m od_ 120 211 624 450 0_0. docx @ 11 4 @ @ 1

Pos: 160 /01 0 Univ ersal mod ule/L eerz eile @ 0\m od_ 120 211 624 450 0_0. docx @ 11 4 @ @ 1

Pos: 161 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 70 Stö run gen und Fe hlers uche /07 0-0 030 Kont akt im S ervice fall_L abo r-RVC @ 25\ mod _14 049 8406 875 8_6 8.do cx @ 1 842 79 @ 2 @ 1

7.3 Service contact

In the event of queries, malfunctions, or spare part enquiries:

From Germany:
Contact
Martin Christ Gefriertrocknungsanlagen GmbH
An der Unteren Söse 50
37520 Osterode (Germany)
Tel. +49 (0) 55 22 / 50 07-44 44
E-mail: [email protected]

Outside Germany:
Contact our agency in your country. All agencies are listed at
www.martinchrist.de [Sales Partners]

NOTE

If you would like to utilise our after-sales-service, please state the type of
your freeze-dryer and its serial number.

Pos: 162 /01 0 Univ ersal mod ule/Absc hnittsw echsel @ 0\ mod _12 021 245 140 62_ 0.docx @ 4 18 @ @ 1

Pos: 163 /01 0 Univ ersal mod ule/Seit enwec hsel @ 0\m od_ 120 211 624 4312 _0. docx @ 10 5 @ @ 1