Error message
Possible cause
Remedy
Ice condenser temperature 20°C
The ice condenser temperature rises to
a non-permissible value during the
freeze-drying process. The pressure
control valve will close automatically in
order to prevent the vacuum pump from
being damaged.
• Insufficient cooling.
• Observe the behaviour of the
refrigeration system. If
necessary, inform the service
department (see chapter 7.3 -
"Service contact").
Defective ice condenser temperature
sensor
• The temperature sensor is not
connected or it is defective.
• Inform the service department
(see chapter 7.3 - "Service
contact").
Defective temperature sensor of the
defrosting heater
• The temperature sensor is not
connected or it is defective.
• Inform the service department
(see chapter 7.3 - "Service
contact").
Mains power failure (see chapter 7.1.1
- "Power failure")
• The power supply has been
interrupted.
• The mains power connector is
not connected.
• The fuses have tripped.
• The mains power switch has
been switched off.
• Check the mains fuse.
• Connect the mains power
connector firmly.
• Check the customer-provided
fuses.
• Switch the mains power switch
on.
Default settings loaded
All of the settings have been reset to the
delivery state of the freeze-dryer.
• Error in the parameter memory.
• Manual execution in the service
menu.
• No measure required.
IO communication error (0x20)
• Error in the LDplus IO module.
• Inform the service department
(see chapter 7.3 - "Service
contact").
Pos: 159 /01 0 Univ ersal mod ule/L eerz eile @ 0\m od_ 120 211 624 450 0_0. docx @ 11 4 @ @ 1
Pos: 160 /01 0 Univ ersal mod ule/L eerz eile @ 0\m od_ 120 211 624 450 0_0. docx @ 11 4 @ @ 1
Pos: 161 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 70 Stö run gen und Fe hlers uche /07 0-0 030 Kont akt im S ervice fall_L abo r-RVC @ 25\ mod _14 049 8406 875 8_6 8.do cx @ 1 842 79 @ 2 @ 1
7.3 Service contact
In the event of queries, malfunctions, or spare part enquiries:
From Germany:
Contact
Martin Christ Gefriertrocknungsanlagen GmbH
An der Unteren Söse 50
37520 Osterode (Germany)
Tel. +49 (0) 55 22 / 50 07-44 44
E-mail: [email protected]
Outside Germany:
Contact our agency in your country. All agencies are listed at
www.martinchrist.de [Sales Partners]
NOTE
If you would like to utilise our after-sales-service, please state the type of
your freeze-dryer and its serial number.
Pos: 162 /01 0 Univ ersal mod ule/Absc hnittsw echsel @ 0\ mod _12 021 245 140 62_ 0.docx @ 4 18 @ @ 1
Pos: 163 /01 0 Univ ersal mod ule/Seit enwec hsel @ 0\m od_ 120 211 624 4312 _0. docx @ 10 5 @ @ 1