EXFO IQS-1500 Power Meters for IQS-500/600 User Manual
Page 85

Calibrating Other Power Meters with the IQS-1500
Power Meters
77
Viewing Calibration Summary
The standard deviation (all wavelengths) of the mean calibration factor
(CF) is calculated using the following equation:
The Single Wavelength tab allows you to see the results for each series of
measurements taken on the DUT on a particular wavelength.
To switch between wavelengths in the tab:
Use the up and down arrows next to the wavelength list. The tab will be
updated accordingly.
If you want to sort the columns using a particular criterion, simply click on
the header of the corresponding columns. A first click will display the list in
alphabetical order, a second click will reverse the order.
Std. Dev.
CF
i
2
(
)
i
∑
1
N
----
CF
i
i
∑
⎝
⎠
⎜
⎟
⎛
⎞
2
–
⎝
⎠
⎜
⎟
⎛
⎞
N 1
–
(
)
----------------------------------------------------------------
⎝
⎠
⎜
⎟
⎜
⎟
⎜
⎟
⎜
⎟
⎜
⎟
⎛
⎞
100
CF
---------
×
=
%
( )
Calibration
factor
Average
power
measured on
IQS-1500
Source
wavelength
Sequential
number of
calibration
factor
measurement
series