Calibrating Other Power Meters with the IQS-1500

72

IQS-1100/IQS-1500

Nulling Offsets for Calibration

Nulling Offsets for Calibration

The Null window allows you to perform an offset nulling on your modules
(both the Calibration Power Meter and the DUT Power Meter). This will
provide you with a “clean” measurement, with no electrical offsets to
impede your calibration process.

To perform an offset nulling on your modules:
1.
Place a protective cap on the detector ports.

2. Confirm that you have used the protective caps by checking the

corresponding box in the window.

3. Click Start NULL Session to start the process.