Entering dut measurements – EXFO IQS-1500 Power Meters for IQS-500/600 User Manual
Page 83

Calibrating Other Power Meters with the IQS-1500
Power Meters
75
Entering DUT Measurements
Entering DUT Measurements
After taking the measurements on your reference power meter, you must
take them on the DUT, in the DUT Meter Measurement window. You can
enter as many measurements as needed.
To add DUT measurements to your calibration procedure:
1. Refer to the DUT display and read the power value.
2. In the DUT Meter Measurement window, enter this value in the New
Value box, then select the proper units using the up and down arrow
buttons.
3. Add the value to the list by clicking .
Note: At this point, the unit selection will be grayed out so as not to change it by
mistake in another entry.
4. Repeat steps 1 through 3 for each measurement you want to add.