Sense[1..n]:power[:dc]:unit – EXFO IQS-1500 Power Meters for IQS-500/600 User Manual
Page 151

SCPI Command Reference
Power Meters
143
Product-Specific Commands—Description
:SENSe[1..n]:POWer[:DC]:UNIT
Description
This command changes the absolute
measurement display unit: dBm or W for the
specified channel.
Syntax
:SENSe[1..n]:POWer[:DC]:UNIT
Parameter(s)
Unit:
The program data syntax for the first parameter is
defined as a
element. The allowed DATA> elements for this parameter are: DBM|W. Power unit in dBm or Watt, where: DBM - sets the power display to dBm W | WATT - sets the power display to Watt (pW,nW,uW,Mw…) Example(s) SENS:POW:UNIT W Notes This command is not executed if a data acquisition is in progress. In that case, the "Acquisition in progress" message will be returned. See Also SENSe:POWer[:DC]:UNIT? UNIT:POWer UNIT:POWer?