C calibration report sample, Power meters – EXFO IQS-1500 Power Meters for IQS-500/600 User Manual
Page 165

Power Meters
157
C Calibration Report Sample
Here is an example of long-form report. The short-form report will be
similar, but contains less uncertainty-related information.
1 of 2
CALIBRATION REPORT
Before the calibration was performed, the instrumentation was allowed to reach equilibrium with the surrounding test
environment. The nullings were performed on the reference standard and on the DUT as specified by the manufacturer. The
Calibration conditions are listed below.
The reference standard (calibration power meter) is traceable to NIST.
The Expanded uncertainties are expressed at approximately a level of confidence of 95 % using a coverage factor of k = 2.
The DUT readings must be multiplied by the Calibration factor to have the correct value.
EXFO Inc.
465 Godin Ave
Vanier (QC) Canada
G1M 3G7
Customer: EXFO
Address: Service center, EXFO3,
Description: 4-Ch power meter
Serial no.: EME 0103 2641
Calibration date: 2002-11-27
Calibration software: 1.0.19
Model no.: IQ-1203
Report no.: 2
Calibration conditions
Condition
Value
Temperature
23 oC +/- 1 oC
Relative humidity
50 % +/- 10 %
Source :
IQ-2403 EME 0007 1935
Fiber type
SM
Connector
FC/UPC
Bandwidth
0.1 nm
Wavelength
1551.0
Source :
IQ-2404 EME xxxx xxxx
Fiber type
SM
Connector
FC/UPC
Bandwidth
0.1 nm
Wavelength
1626.0
Standard used to establish traceability
Description
Serial number
Calibration date
Calibration due date
IQS-1502 - Calibration power meter
29219-1K
2002-01-22
2003-01-22
Calibration summary
Dut
wavelength
(nm)
Source
wavelength
(nm)
Mean power
N
Calibration factor
Standard deviation
(%)
Expanded
uncertainty
(%)
1550.0
1551.00
4.824 dBm
2
0.9802
2.87
5.55
1625.0
1626.0
2.076 dBm
2
0.9857
2.07
4.07