C Calibration Report Sample

Here is an example of long-form report. The short-form report will be
similar, but contains less uncertainty-related information.

1 of 2

CALIBRATION REPORT

Before the calibration was performed, the instrumentation was allowed to reach equilibrium with the surrounding test
environment. The nullings were performed on the reference standard and on the DUT as specified by the manufacturer. The
Calibration conditions are listed below.

The reference standard (calibration power meter) is traceable to NIST.

The Expanded uncertainties are expressed at approximately a level of confidence of 95 % using a coverage factor of k = 2.
The DUT readings must be multiplied by the Calibration factor to have the correct value.

EXFO Inc.

465 Godin Ave

Vanier (QC) Canada

G1M 3G7

Customer: EXFO

Address: Service center, EXFO3,

Description: 4-Ch power meter

Serial no.: EME 0103 2641

Calibration date: 2002-11-27

Calibration software: 1.0.19

Model no.: IQ-1203

Report no.: 2

Calibration conditions

Condition

Value

Temperature

23 oC +/- 1 oC

Relative humidity

50 % +/- 10 %

Source :

IQ-2403 EME 0007 1935

Fiber type

SM

Connector

FC/UPC

Bandwidth

0.1 nm

Wavelength

1551.0

Source :

IQ-2404 EME xxxx xxxx

Fiber type

SM

Connector

FC/UPC

Bandwidth

0.1 nm

Wavelength

1626.0

Standard used to establish traceability

Description

Serial number

Calibration date

Calibration due date

IQS-1502 - Calibration power meter

29219-1K

2002-01-22

2003-01-22

Calibration summary

Dut

wavelength

(nm)

Source

wavelength

(nm)

Mean power

N

Calibration factor

Standard deviation

(%)

Expanded

uncertainty

(%)

1550.0

1551.00

4.824 dBm

2

0.9802

2.87

5.55

1625.0

1626.0

2.076 dBm

2

0.9857

2.07

4.07