Calibration report sample – EXFO IQS-1500 Power Meters for IQS-500/600 User Manual
Page 166

Calibration Report Sample
158
IQS-1100/IQS-1500
2 of 2
Comments:
Prepared by:
Approved by:
4-Ch power meter
Model no.: IQ-1203
Serial no.: EME 0103 2641
Calibration date: 2002-11-27
Calibration uncertainties at DUT wavelength = 1550.0 nm
Type A uncertainties (%)
Type B uncertainties (%)
Standard deviation
Type A other
2.87
1.10
Source power stability
Reference meter spectral sensitivity
DUT spectral sensitivity
Setup reproducibility
Connector
Type B other
1.00
1.00
1.00
1.00
1.00
1.00
Reference meter calibration uncertainty (%)
Reference meter
2.00
Calibration uncertainties at DUT wavelength = 1625.0 nm
Type A uncertainties (%)
Type B uncertainties (%)
Standard deviation
2.07
Source power stability
Reference meter spectral sensitivity
DUT spectral sensitivity
Setup reproducibility
0.00
1.00
1.00
1.00
Reference meter calibration uncertainty (%)
Reference meter
2.00
11/27/2002
Metrologist name
Date
Metrologist
Metrology Office
Date