beautypg.com

Viewing calibration summary – EXFO IQS-1500 Power Meters for IQS-500/600 User Manual

Page 84

background image

Calibrating Other Power Meters with the IQS-1500

76

IQS-1100/IQS-1500

Viewing Calibration Summary

The average value for your measurements will automatically be
updated below the New Value box.

If you want to remove a value from the list, click

.

Once you have entered all of the DUT measurements, proceed to the next
step by clicking Next.

Viewing Calibration Summary

Once your calibration is complete, you can view a summary of the results
in the Summary window. This is also where you can manage the results
you have obtained. You can delete erroneous results, or go back to a
previous step to obtain more measurements.

The calibration factor (CF) is calculated using the following equation:

where

P

1500

= IQS-1500 measured power

P

DUT

= DUT measured power

CF

P

1500

W

( )

P

DUT

W

( )

------------------------

=