Viewing calibration summary – EXFO IQS-1500 Power Meters for IQS-500/600 User Manual
Page 84

Calibrating Other Power Meters with the IQS-1500
76
IQS-1100/IQS-1500
Viewing Calibration Summary
The average value for your measurements will automatically be
updated below the New Value box.
If you want to remove a value from the list, click
.
Once you have entered all of the DUT measurements, proceed to the next
step by clicking Next.
Viewing Calibration Summary
Once your calibration is complete, you can view a summary of the results
in the Summary window. This is also where you can manage the results
you have obtained. You can delete erroneous results, or go back to a
previous step to obtain more measurements.
The calibration factor (CF) is calculated using the following equation:
where
P
1500
= IQS-1500 measured power
P
DUT
= DUT measured power
CF
P
1500
W
( )
P
DUT
W
( )
------------------------
⎝
⎠
⎛
⎞
=