beautypg.com

EXFO IQS-1500 Power Meters for IQS-500/600 User Manual

Page 139

background image

SCPI Command Reference

Power Meters

131

Product-Specific Commands—Description

:SENSe[1..n]:CORRection:COLLect:

ZERO:ASYNchronous:ALL

Description

This command performs an asynchronous offset

nulling measurement on all channels.

Syntax

:SENSe[1..n]:CORRection:COLLect:ZERO:ASYNc

hronous:ALL

Parameter(s)

None

Example(s)

SENS:CORR:COLL:ZERO:ASYN:ALL

Notes

This command is not executed if a data

acquisition is in progress. In that case, the

"acquisition in progress" message will be

returned. This command will take at least 30

seconds to complete.

See Also

SENSe:CORRection:COLLect:ZERO

SENSe:CORRection:COLLect:ZERO:ALL

SENSe:CORRection:COLLect:ZERO:ASYNchrono

us

This manual is related to the following products: