Technical Specifications

106

IQS-1100/IQS-1500

GENERAL SPECIFICATIONS

Size (H

× W × D)

12.5 cm

× 3.6 cm × 28.2 cm

(4

15

/

16

in

× 1

7

/

16

in

× 11

1

/

8

in)

Weight

0.63 kg

(1.4 lb)

Temperature

operating

23

°C ± 5 °C (73

°F ± 9 °F)

storage

—40

°C to 70 °C

(—40

°F to 158 °F)

Relative humidity

0 % to 80 % non-condensing

Model

IQS-1502-Q0

IQS-1502-Q1

Detector type

Ge

Ge

Detector size (mm)

5

5

Power range

1

(dBm)

+5 to —60

+5 to —60

Wavelength range (nm)

750 to 1800

750 to 1800

Linearity

1,2

(dB)

± 0.01

± 0.01

(+5 dBm to —50 dBm)

(+5 dBm to —50 dBm)

Uncertainty

6

± 2 %

3

≤ ± 0.9 %

4

at 3

λ

(at 1310.0 ± 0.1 nm and 1550.0 ± 0.1 nm)

(—10 dBm CW)

(0 dBm to —10 dBm CW)

± 3 %

3

± 3 %

4

(750 nm to 1700 nm)

(750 nm to 1700 nm)

(—30 dBm to —40 dBm CW)

(—30 dBm to —40 dBm CW)

Power resolution (dB)

0.001

0.001

Wavelength resolution (nm)

0.1

0.1

Applicable fiber type (μm)

9/125 (B); 50/125 (C); 62.5/125 (D)

9/125 (B); 50/125 (C); 62.5/125 (D)

Connector type

5

B: 89

B: 58; 88; 89; 90; 91
C/D: 50; 54; 74

IQS-150

X

-

XX

-

X

-

XX

ORDERING INFORMATION

Model
IQS-150

2

-

Q0

-

B

= 5 mm TEC-Ge detector, 2 % uncertainty at (1310 ± 0.1) nm and (1550 ± 0.1) nm, Singlemode 9/125 μm

7

IQS-150

2

-

Q1

-

B

= 5 mm TEC-Ge detector, 0.9 % uncertainty calibrated at NIST, Singlemode 9/125 μm

IQS-150

2

-

Q1

-

C

= 5 mm TEC-Ge detector, 0.9 % uncertainty calibrated at NIST, Multimode 50/125 μm

IQS-150

2

-

Q1

-

D

= 5 mm TEC-Ge detector, 0.9 % uncertainty calibrated at NIST, Multimode 62.5/125 μm

Test jumper connector

50

= FC/PC

9

54

= SC/PC

9

58

= FC/APC narrow key

8

Example: IQS-1502-Q1-B-89

74

= ST/PC

9

88

= SC/APC

8

89

= FC/UPC narrow key

8

90

= ST/UPC

8

91

= SC/UPC

8

STANDARD ACCESSORIES

User guide, fiber-optic adapter
(FOA), one reference test jumper
and Certificate of Calibration

Notes
1. At 1310 nm and 1550 nm.
2. At a ± 1 °C constant temperature.
3. Q0 option, uncertainty at EXFO

reference conditions:

• ± 2 % uncertainty at (1310.0 ± 0.1) nm

and (1550.0 ± 0.1) nm, with 9/125 μm
(B) fiber, and source spectral width
(FWHM) < 10 nm.

• ± 3 % uncertainty from 750 nm to

1700 nm, with 62.5/125 μm (D) fiber,
and source spectral width (FWHM)
≤ 12 nm.

• FC/UPC connector (ceramic ferrule)

with FOA-322.

4. Q1 option, calibration at NIST
a) •

≤ ± 0.9 % uncertainty at three
user-specified wavelengths within
the following ranges at NIST
reference conditions:

786 nm ± 0.5 nm
(840 to 860) nm ± 0.25 nm
(1280 to 1330) nm ± 0.25 nm
(1520 to 1625) nm ± 0.25 nm

• 9/125 to 62.5/125 μm (B to D) fiber.
• FC, ST or SC connector.
• Fiber-optic adapter (FOA) used

at NIST.

• Source spectral width (FWHM)

< 10 nm.

b) ± 3 % uncertainty from 750 nm to

1700 nm, 62.5/125 μm (D) fiber,
source spectral width (FWHM)
< 12 nm, and FC/UPC connector
(ceramic ferrule) with FOA-322.

5. See Ordering Information for details.
6. All uncertainties are valid on the

day of calibration after a warm-up
time of 20 minutes, and specified
with a 95 % confidence level.

7. Test jumper FC/UPC only
8. Singlemode only
9. Multimode only