Calibrating Other Power Meters with the IQS-1500

Power Meters

69

Entering Calibration Uncertainties

The total uncertainty is calculated by adding (using the root of the sum of
the squares) the Type B, Type A and Reference Meter uncertainties and
multiplying the results by a factor of two to obtain a confidence level of
95 %, as shown in the following equation:

where

U

t

= total uncertainty

U

Bi

= Type B uncertainties

U

Ai

= Type A uncertainties

U

1500

= IQS-1500 calibration uncertainty

The mandatory Type B uncertainties are:

³

Source power stability

³

Reference Meter spectral sensitivity

³

DUT spectral sensitivity

³

Setup reproducibility

U

t

2

U

2

Bi

3

-----------

i

U

2

Ai

N

------------

i

U

1500

2

-------------