beautypg.com

A&D Weighing HW-1200KGV3 User Manual

Page 70

background image

Page 68

HV-G/HW-G Series

17. Options

17.3. RS-422/ RS-485 / Relay Output (OP-04)

Replacing with the standard RS-232C interface, the RS-422/ RS-485 interface can

connect up to 16 scales and control them from a computer.


The RS-422/ RS-485 interface has the following two modes.

Stream mode

Outputs data continuously.

Command mode Controls the scale using commands from a computer.


To use the RS-422/ RS-485 interface, set the parameters of the "Baud rate (

f4

)",

“Address (

f11

)", "Output mode (

f5

)" and " Format (

f15

)" in the function table.


Transmission system

EIA RS-422/ RS-485

Transmission

Asynchronous, bi-directional, half-duplex

Data format

Baud rate

2400, 4800, 9600 bps

Data

7

bits

Parity

1 bit, Even

Start bit

1 bit

Stop bit

1 bit

Code

ASCII

Terminator

CR LF (CR: 0Dh, LF: 0Ah)

Address

01 to 99

Address parameter (

f11

) of function table.


Relay

Solid-state-relay

Maximum voltage

50V DC

Maximum current

100mA DC

Maximum resistance

8

Ω


Adaptable connector

TM:BLA9 (of accessory)


Pin

connections

Pin No. Signal name Direction

Description

A

1

SDA

Output

RS-422/485 transmission A terminal

A

2

SDB

Output

RS-422/485 transmission B terminal

A

3

RDA

Input

RS-422/485 receive A terminal

A

4

RDB

Input

RS-422/485 receive B terminal

A

5 TRM

-

120

Ω terminator

A

6

HI

Output

Relay output of HI

A

7

OK

Output

Relay output of OK

A

8

LO

Output

Relay output of LO

A

9

COM

-

Relay common terminal

A