Stream mode, Preparation and performance (examples) – A&D Weighing HW-1200KGV3 User Manual
Page 61

HV-G/HW-G Series
Page 59
16. RS-232C Serial Interface
16.2. Stream Mode
The scale outputs the current weighing data at the time the display is refreshed.
The scale does not output data while in the setting mode.
Averaging of function table
Refresh rate
A
f13 0 Fast response
Sensitive value
Approximately 7 times/s while the display is unstable,
Approximately 4 times/s when the display is stable
A
f13 1
Ú
A
f
f13 2 Slow response
Stable
value
Approximately 4 times/s
A
16.2.1. Preparation and Performance
(Examples)
Step 1 Turn off the display.
Press and hold the
ZERO
key and press the
ON/OFF
key. The function table is displayed.
Step 2 Press the
ENTER
key to display f4 ("Baud rate").
Select a parameter of "Baud rate" with the
∧
key.
Step 3 Press the
ENTER
key to display f5 ("Output
mode").
Select "Stream mode ( f5 0 )" with the
∧
key.
Step 4 Press the
ENTER
key to display f15 ("Format").
Select "Format 1 ( f15 0 )" with the
∧
key.
Step 5 Press the
ENTER
key to display end .
Step 6 Press the
ENTER
key to store the new parameters.
The scale returns to the normal weighing mode and
outputs the data continuously.