Operation and performance (examples) – A&D Weighing HW-1200KGV3 User Manual
Page 41
HV-G/HW-G Series
Page 39
11. Upper/Lower Comparator Function
11.2. Operation and Performance (Examples)
Example 1
This example is set as follows:
Function table
f6 1
(The scale always compares the weight value
even when the value is nearly-zero.)
Upper limit value (HI)
7.000 kg
Lower limit value (LO)
6.500 kg
Case
The comparison starts when the scale is turned on.
When the current value is less than 6.500kg, LO is displayed.
When the current value is between 6.500kg and 7.000kg, OK is displayed.
When the current value is greater than 7.000kg, HI is displayed.
Example 2
This example is set as follows:
Function table
f6 4
(When the
SET
key is pressed and the stability
mark is displayed, the scale compares the
current weight value with the upper limit value
and the lower limit value immediately.)
Upper limit value (HI)
2.000 kg
Lower limit value (LO)
-1.000 kg
Case
The comparison starts when the
SET
key is pressed and the stability mark is
displayed.
When the current value is less than -1.000kg, LO is displayed.
When the current value is between -1.000kg and 2.000kg, OK is displayed.
When the current value is greater than 2.000kg, HI is displayed.