Operation and performance (examples) – A&D Weighing HW-1200KGV3 User Manual
Page 48
Page 46
HV-G/HW-G Series
13. Simple Batch Function
Step 6 Press the
MODE
key to display the blinking HI (of the
final value).
Step 7 Set the final value using the following keys.
<
key
Selecting the figure to be changed.
∧
key
Changing the number of the figure.
Step 8 Press the
ENTER
key to store the new parameter and
display the blinking OK (of preliminary value).
Step 9 Set the preliminary value using the following keys.
<
key
Selecting the figure to be changed.
∧
key
Changing the number of the figure.
Step 10 Press the
ENTER
key to store the new parameter and
display the blinking LO (of zero band).
Step 11 Set the zero band using the following keys.
<
key
Selecting the figure to be changed.
∧
key
Changing the number of the figure.
Step 12 Press the
ENTER
key to store the new parameter and
return to the normal weighing mode.
13.2. Operation and Performance (Examples)
Step 1 Select the parameter
f6 8
of the function table.
Step 2 Set the parameters of the final value, preliminary value and zero band.
Step 3 The comparison result is always displayed.
Comparison Condition
Gross < Zero band ..........LO is displayed and output.
Final - Preliminary ≦ Net.....................OK is displayed and output.
Final
≦
Net.....................OK, HI are displayed and output.