A&D Weighing HW-1200KGV3 User Manual
Page 57

HV-G/HW-G Series
Page 55
15. The Function Table
Item
Display Description
A
Selects the method to hold the current weight value. When the
weight value becomes nearly-zero or changes more than 25%
+30 digits, hold display is canceled.
A
A
A
f12 0 # Not used
A
f12 1
The hold function is ON or OFF alternately by the
F
key.
A
Hold
f12 2
When the stability mark is displayed, the display is held.
A
Selects readability and response.
A
f13 0 # Fast response, sensitive value
A
f13 1
Ú
A
Averaging
f13 2
Slow response, stable value
A
Selects a storable minimum unit mass in the counting mode.
A
f14 0 # Stores a unit mass in the unit of a digit.
A
f14 1
Stores a unit mass in the unit of one eighth digit.
A
Precision
of unit
mass
f14 2
Stores a unit mass, if the total of sample mass is greater than 5
digits.
A
A
Refer to "16.1. Data Format".
A
f15 0 # Format 1 (A&D general format for scales, balances)
A
f15 1 Format 2 (Older HV-A/ HW-A format)
A
f15 2
Not used
Format
f15 3
Format 1 (Return command is available for all commands.)
Used when Type V scale is equipped with option OP-06, a
built-in impact dot matrix printer.
f16 0
f16 1 #
to
Number of
carriage
return for
the printer
f16 9
Selects the number of carriage return after the data is printed.
Option OP-06 for type V is required.
Use Printout of date. Caution Date is not updated.
Printout of the management number for the scale.
A
f17
000000 #
The function number “f17” is not displayed.
No printout
A
A
Date
others
The relation between printout and display is as follows:
Display 990721
Printout
99/07/21
A
A
"#": factory settings
The "nearly-zero" is within ±5 digits from zero point in the unit of kg.
The “digit” is a unit of display, and is equivalent to the minimum measurable mass.