Complete calibration procedure, Gravity acceleration correction, Preparation – A&D Weighing HW-1200KGV3 User Manual
Page 51

HV-G/HW-G Series
Page 49
14. Calibration (Adjusting Scale)
14.2. Complete Calibration Procedure
14.2.1. Gravity Acceleration Correction
Step 1 Turn on the display.
Open the rear cover of the display unit. Locate the
CAL switch inside.
Press and hold the
CAL
switch to enter the calibration
mode.
Then
Cal 0
is displayed.
Step 2 Press the
ZERO
key to enter the gravity acceleration
correction mode.
Step 3 Set your local gravity acceleration using the following
keys.
<
key
Selecting the figure to be changed.
∧
key
Changing the number of the figure.
Step 4 Press the
ENTER
key to store the new value.
Then
Cal 0
is displayed.
Step 5 Press the
CAL
switch again. Then the scale returns to
the normal weighing mode.
14.2.2. Preparation
Step 6 Confirm the environmental conditions as follows:
Maintain a constant temperature and stable power.
Install the scale on a solid floor where there is no draft, vibration, strong magetic
fields or direct sunlight.
Refer to "4.1. Installing the Scale".
Step 7 Keep the display turned on for at least 30 minutes to warm up the scale.