beautypg.com

Counting mode, Storing a unit mass, Caution – A&D Weighing HW-1200KGV3 User Manual

Page 31

background image

HV-G/HW-G Series

Page 29

8. Counting Mode

8. Counting Mode

The counting mode is the function to convert the total mass value (total weight) of

articles to a count, when each article has the same mass value.

To use this function, store a unit mass in advance.
Even if the AC adapter or the batteries is removed, the unit mass is maintained in

non-volatile memory.

8.1. Storing a Unit Mass

Step 1 Press the

MODE

key to display the unit .



Step 2 Press the

SET

key to enter the mode that stores a

unit mass.



Step 3 Press the

key to select the number of samples.

The greater the quantity of samples, the greater the
accuracy of the count.

5 pieces, 10 pieces, 20 pieces, 50 pieces,

100 pieces


Step 4 Place the container on the pan.

Press the

TARE

key.


Step 5 Place the number of samples selected at step 3.

Wait for the stability mark to be displayed. Press the

ENTER

key to store. The count is displayed.

Caution

When the sample total mass value is too small and
it is not possible to calculate a unit mass, the scale
displays

lo ut

and returns to step 3. Increase the

number of samples. More than 5 digits of the total
sample mass, in the unit of kg, are required.

When the unit mass is too light to store, the scale
displays

lo ut .

In this case, the unit mass will not be

stored even if the number of samples is increased.

Pressing

the

ENTER

key, after

lo ut

is

displayed, displays the next unit.

Step 6 Remove the samples and the container from the pan.