10. Accumulation Function
10. Accumulation Function
This
function
counts
the number of times articles are weighed, calculates the total mass
value and can display the number (accumulation count) and accumulated mass value.
The accumulation function is displayed with up to 6 digits. The balance can not display 7 or
more digits, therefore the leading digits are not displayed.
Example: With 60K type, when importing the data of 17 accumulations of up to 60kg (60.000
X 17=1020.000), the balance displays this as “020.000”.
To use this function, set the parameters of the "Accumulation function (
f8
)" in the
function table in advance.
To use the built-in printer, set the parameters of the "Print mode (
f9 5
)" in the
function table in advance.
The accumulation count and accumulated mass value are stored in the scale even if
the power is removed.
Operation and Keys
The display of the accumulation count has a blinking or
without a unit.
The display of the accumulated mass value has a unit and a blinking or
.
Press
the
MODE
key to display the accumulation count and accumulated mass value.
Press
the
ZERO
key in the accumulation function (with a blinking or
) to reset
the current function (The count and accumulated mass value become zero.)
When the optional built-in printer is installed and the
key is pressed, the
accumulation data, date and data number are printed. Date is set at function table f17 .
Caution
The accumulation function is available only when weighing is performed in the
same unit.
Parameter List and Word Definition
The "nearly-zero" is within ±5 digits from the zero point in the unit of kg.
The "digit" is a unit of display, and is equivalent to the minimum measurable mass.
The "zero point" is the fundamental starting point to weigh anything.
Function table
Description
A
f8 0
Accumulation function not used.
A
f8 1
The scale accumulates the data, if the
F
key is pressed, when the
display is a positive stable value and not nearly-zero. The next
accumulation can be performed after the display becomes
nearly-zero or a negative value.
A
A
A
A
f8 2
The scale accumulates the data, if the
F
key is pressed, when the
display is a stable value and not nearly-zero. The next accumulation
can be performed after the display becomes nearly-zero.
A
A
A