Parameter list – A&D Weighing HW-1200KGV3 User Manual
Page 54
Page 52
HV-G/HW-G Series
15. The Function Table
15.2. Parameter List
Item
Display Description
A
The type L scale is turned off after 5 minutes of no operation,
when zero is displayed.
A
A
f1 0 # Not
used
A
Automatic
turning off
f1 1
Used
A
Selects how to change the weighing range for HV-L and HV-G
series.
A
f2 0 # Automatic
range
A
Weighing
range
f2 1
Manual range
A
Selects the first unit at the time when the scale turns on.
A
f3 0 # Kg
A
f3 1
Lb
A
f3 2
Oz
A
Unit
f3 3
lb-oz
A
Selects the transmission rate of the serial interface (RS-232C/ 422/485).
A
f4 0 # 2400
bps
A
f4 1
4800 bps
A
Baud rate
f4 2
9600 bps
A
Selects the mode for the serial interface (RS-232C/ 422/485).
A
f5 0 # Stream mode. (Refer to "16.2. Stream Mode")
A
f5 1
Command mode. (Refer to "16.3. Command Mode")
A
f5 2
A data is output, when the
key is pressed.
A
f5 3
Auto-print +
When the display becomes a positive stable value and not
nearly-zero, the scale outputs the data automatically. Next
output can be performed after the display becomes
nearly-zero or a negative value.
A
A
A
A
f5 4
Auto-print +/-
When the display becomes a stable value and not
nearly-zero, the scale outputs the data automatically. Next
output can be performed after the display becomes
nearly-zero.
A
A
A
A
Output
mode
f5 5
At each end of full/dribble batch function, the data is output.
A
Selects the comparator mode.
A
f6 0 #
Press the
SET
key to always compare the current weight value.
Press the
SET
key again to stop the comparison.
A
A
f6 1
The scale always compares the weight value.
A
f6 2
Press the
SET
key to always compare the weight value when not
nearly-zero. Press the
SET
key again to stop the comparison.
A
A
Comparator
f6 3
The scale always compares the weight value when not
nearly-zero.
A
"#" :factory settings. The "nearly-zero" is within ±5 digits from zero point in the unit of kg.