Step height measurement – KEYENCE TM-3000 Series User Manual
Page 5

5
Select OUT2 on screen 2 and make the setting for measuring (2) the step height.
Sliding the [PROG/RUN] switch on the side of the console switches from the setting screen to the
measurement screen.
The measurement results for (1) Diameter and (2) Step Height are displayed.
To set scaling or tolerance for measurement values, see page11 in this manual.
Select “OUT2” on screen 2 on the previous page.
Select “Step Height” for “Measurement type” on screen 4.
Select “Area 1/Area 2” on screen 4 and set the area as
shown on screen 4.
For area operation, see page 8 in this manual.
Press the [ESCAPE] key on the console to return to screen 1.
The step height measurement setting (2) is complete.
* To measure correcting the tilt, see the position correction function (page 8 in this manual).
Pressing the [SCREEN] key
switches the screen according
to each setting.
Screen 4
III -
22
Manual
P5-48
Step height measurement
XXX
Measuring the step height of a target
Setting procedure
Checking the measurement status
[SCREEN]
8
9
7
8
9
10