Measurement – KEYENCE TM-3000 Series User Manual
Page 4

4
III -
11
Manual
P5-43
Make the detailed settings for each OUT.
Measure the diameter for OUT1 and step height for OUT2 respectively.
Perform the master registration beforehand for this measurement.
For setting the master registration, see page 6 in this manual.
Slide the [PROG/RUN] switch on the side of the console to
display the settings screen (screen 1).
The screen switches from the measurement screen to the setting
screen.
Select “Measure set” on screen 1.
Press the [ENTER] key on the console to select the item.
Select “OUT1” on screen 2.
Select “Diameter” in “Measurement type” on screen 3
Select “Area 1” and set the area vertically long.
For area operation, see page 8 in this manual.
Confirm the “Measured content” is set to “Average” and “Tilt
corr” is set to “ON”.
* Setting the tilt correction (Tilt corr) to “ON” enables the diameter to
be measured correctly even if the target has a tilt.
Press the [ESCAPE] key on the console to return to screen 2.
The diameter measurement setting (1) is complete.
This measures the stepped shaft.
Diameter measurement
XXX
Measuring the outer diameter of a target
(1) Diameter
measurement
(2) Step height measurement
Setting procedure
Screen 2
Screen 3
Screen 1