8 Troubleshooting
62
B-290 Operation Manual, Version I
8
Troubleshooting
This chapter helps to resume operation after a minor problem has occurred with the instrument. It lists
possible occurrences, their probable cause and suggests how to remedy the problem.
The troubleshooting table below lists possible malfunctions and errors of the instrument. The operator
is enabled to correct some of those problems or errors by him/herself. For this, appropriate corrective
measures are listed in the column “Corrective measure”.
The elimination of more complicated malfunctions or errors is usually performed by a BUCHI technical
engineer who has access to the official service manuals. In this case, please refer to your local BUCHI
customer service agent.
8 .1
Error messages and their remedy
Table 8-1: Error messages and their remedy
Code
Description
Possible cause
Corrective measure
001
Entry temperature sensor
interruption
Defective sensor, sensor cable or
internal wiring
Contact the BUCHI customer
service
002
Entry temperature sensor
short-circuit
Defective sensor, sensor cable or
internal wiring
Contact the BUCHI customer
service
003
Entry temperature exces-
sively high (> 230 °C)
Defective heating control or heating
relay
Contact the BUCHI customer
service
004
Exit temperature sensor
interruption
Defective sensor, sensor cable or
internal wiring
Contact the BUCHI customer
service
005
Exit temperature sensor
short-circuit
Defective sensor, sensor cable or
internal wiring
Contact the BUCHI customer
service
006
Exit temperature exces-
sively high
Defective sensor, sensor cable or
internal wiring
Contact the BUCHI customer
service
010
Malfunction of frequency
converter
Defective frequency converter or wiring
Contact the BUCHI customer
service
011
Heating malfunction
See “System does not heat up”
See “System does not heat up”
Only for operation with the Remote Control Panel
100
Connection to Remote
Control Panel interrupted or
disconnected
Defective connection cable or instru-
ment is not plugged in
Check connection cable for
damage and insert plug, if neces-
sary
Only for operation with the Inert Loop B-295
200
Connection to B-295 inter-
rupted or disconnected
Defective connection cable or instru-
ment is not plugged in
Check connection cable for
damage and insert plug, if neces-
sary
201
B-295 plugged in while
operation in process
Connection cable plugged in after
switching on instrument
Restart the system
202
O
2
alarm not given during
start-up
Oxygen sensor “spent”
Oxygen measuring equipment defective
Replace the oxygen sensor
Contact the BUCHI customer
service