8 Troubleshooting

62

B-290 Operation Manual, Version I

8

Troubleshooting

This chapter helps to resume operation after a minor problem has occurred with the instrument. It lists
possible occurrences, their probable cause and suggests how to remedy the problem.
The troubleshooting table below lists possible malfunctions and errors of the instrument. The operator
is enabled to correct some of those problems or errors by him/herself. For this, appropriate corrective
measures are listed in the column “Corrective measure”.
The elimination of more complicated malfunctions or errors is usually performed by a BUCHI technical
engineer who has access to the official service manuals. In this case, please refer to your local BUCHI
customer service agent.

8 .1

Error messages and their remedy

Table 8-1: Error messages and their remedy

Code

Description

Possible cause

Corrective measure

001

Entry temperature sensor
interruption

Defective sensor, sensor cable or
internal wiring

Contact the BUCHI customer
service

002

Entry temperature sensor
short-circuit

Defective sensor, sensor cable or
internal wiring

Contact the BUCHI customer
service

003

Entry temperature exces-
sively high (> 230 °C)

Defective heating control or heating
relay

Contact the BUCHI customer
service

004

Exit temperature sensor
interruption

Defective sensor, sensor cable or
internal wiring

Contact the BUCHI customer
service

005

Exit temperature sensor
short-circuit

Defective sensor, sensor cable or
internal wiring

Contact the BUCHI customer
service

006

Exit temperature exces-
sively high

Defective sensor, sensor cable or
internal wiring

Contact the BUCHI customer
service

010

Malfunction of frequency
converter

Defective frequency converter or wiring

Contact the BUCHI customer
service

011

Heating malfunction

See “System does not heat up”

See “System does not heat up”

Only for operation with the Remote Control Panel
100

Connection to Remote
Control Panel interrupted or
disconnected

Defective connection cable or instru-
ment is not plugged in

Check connection cable for
damage and insert plug, if neces-
sary

Only for operation with the Inert Loop B-295
200

Connection to B-295 inter-
rupted or disconnected

Defective connection cable or instru-
ment is not plugged in

Check connection cable for
damage and insert plug, if neces-
sary

201

B-295 plugged in while
operation in process

Connection cable plugged in after
switching on instrument

Restart the system

202

O

2

alarm not given during

start-up

Oxygen sensor “spent”
Oxygen measuring equipment defective

Replace the oxygen sensor
Contact the BUCHI customer
service