beautypg.com

5 .10 quick hose couplings, Danger – BUCHI Inert Loop B-295 User Manual

Page 39

background image

5 Putting into operation

39

B-290 Operation Manual, Version I

5 .10 Quick hose couplings

Use
For fast and reliable connecting/disconnecting of polypress tubes e.g. the tubing of Inert Loop B-295
and Dehumidifier B-296.

Installation
• Cut the polypress tube with a sharp knife or a special tube cutter into two halfs with a clean cut.

The tube ends must be straight!

• Slip one hose clamp over each tube end.
• Use one male and one female coupling adapter and install them at the tube ends.
• Secure the coupling adapters with the hose clamps.
• Test all tubes and connectors for tightness prior use!

Female coupling (including a flat seal)

Male coupling

Fig. 5.17: Quick hose couplings

5 .11 Installing the Inert Loop B-295

!

Danger

Death or serious poisoning by gases or particles due to O

2

– sensor or filter malfunction

• Exchange defective O

2

– sensor immediately

• Exchange O

2

– sensor regularly within the specified maintenance intervals

• Exchange clogged filters immediately
• Exchange filters regularly within the specified maintenance intervals
• Dispose of clogged or defective filters immediately

This manual is related to the following products: